To achieve a systematic error better than 0.5% in measuring absorption lines using a push-broom InGaAs imaging spectrometer, it's crucial to develop a comprehensive calibration pipeline. The pipeline should start with a physically robust mathematical model that describes how light interacts with the instrument and how signal and noise propagate through the system.
Mathematical Model of Light and Signal Propagation
Let's denote the input spectral radiance as (L(\lambda)) (in photons per unit area per unit time per unit wavelength). The number of photons collected by a pixel at position ((x, y)) on the detector, corresponding to a specific wavelength (\lambda), can be modeled as:
[N_{ph}(x, y, \lambda) = L(\lambda) \cdot A(\lambda) \cdot \Delta t \cdot QE(x, y, \lambda)]
where (A(\lambda)) is the effective area of the collecting optics, (\Delta t) is the exposure time, and (QE(x, y, \lambda)) is the quantum efficiency of the pixel at ((x, y)) for wavelength (\lambda).
The detected signal in Digital Numbers (DN) can be expressed as:
[S(x, y) = G(x, y) \cdot N_{ph}(x, y, \lambda) + B(x, y) + NL(x, y, S)]
where (G(x, y)) is the gain (in DN/e(^-)), (B(x, y)) is the bias or pedestal, and (NL(x, y, S)) represents non-linearity effects.
Noise in the detected signal includes photon noise, read noise, and other sources. The total noise variance can be approximated as:
[\sigma^2(x, y) = G(x, y)^2 \cdot N_{ph}(x, y, \lambda) + \sigma_{read}^2(x, y) + \sigma_{other}^2]
Calibration Pipeline
-
Bias/Pedestal Removal: Subtract (B(x, y)) from (S(x, y)). This step should be done first as it corrects for the offset.
-
Pixel-Level Non-Linearity Correction: Correct (S(x, y)) for non-linearity effects, (NL(x, y, S)), to obtain a linearized signal.
-
Dark Subtraction: Although the "vignette" detector rows are mentioned to be dark but exhibit an additive signal scaling with exposure time, a proper dark subtraction involves characterizing this dark signal and subtracting it. However, given the description, it seems this step might be implicitly handled in the bias removal if the bias is characterized with dark frames or if the additive signal is considered part of the bias.
-
Stray-Light Subtraction: This step involves estimating and removing stray light. A practical model could involve assuming that stray light is a function of the overall illumination level and can be estimated from the "vignette" rows or other areas not directly illuminated. The placement of this step after non-linearity correction and dark subtraction is justified because stray light estimation might depend on the linearity and dark level of the signal.
-
Gain (DN→e⁻) Conversion: Apply (G(x, y)) to convert (S(x, y)) from DN to electrons. This step is crucial for accurate photon noise estimation.
-
Flat-Field/QE Correction: Correct for (QE(x, y, \lambda)) variations. Given the strong QE roll-off, it's essential to correct for QE per pixel or at least per column if the variation within a column is negligible.
Stray-Light Estimation and Removal
A practical model for stray-light estimation could involve:
- Using the "vignette" detector rows to estimate stray light as a function of exposure time or overall illumination level.
- Assuming stray light is relatively uniform or follows a simple spatial pattern across the detector.
Gain Determination
Given the strong QE roll-off from 1600 nm to 1700 nm, it's crucial to determine gain per pixel or at least per spectral channel (column) if the gain variation within a column is negligible. This is because the QE variation with wavelength affects the signal, and gain correction is necessary for accurate conversion to electrons.
Impact of Omitting Gain Variation or Non-Linearity Correction
Omitting gain variation correction would lead to incorrect conversion of DN to electrons, affecting both flux calibration and noise propagation. Non-linearity correction omission would result in incorrect signal values, especially at higher signal levels, biasing both the measured flux and the noise estimation.
Achieving Systematic Error < 0.5%
To achieve a systematic error better than 0.5%:
- Ensure that the calibration steps are performed with high accuracy.
- Characterize the instrument thoroughly, including non-linearity, gain, QE, and stray light.
- Validate the calibration pipeline with known spectral sources to assess the accuracy of absorption line measurements.
By following this pipeline and ensuring that each step is accurately characterized and applied, it's possible to minimize systematic errors and achieve the desired level of accuracy in measuring absorption lines.