To achieve a systematic error better than 0.5% in measuring accurate absorption lines using a push-broom InGaAs imaging spectrometer, we need to meticulously design a calibration pipeline that accounts for all significant sources of error and noise. Given the steep QE roll-off and the presence of non-uniformities in the detector, the calibration process must be both thorough and precise.
Physical Model of Signal and Noise Propagation
Let's start with a physical model of the signal and noise propagation through the spectrometer:
-
Incident Photon Flux: The number of photons incident on the detector at wavelength (\lambda) and pixel (i) is given by:
[
P_{i}(\lambda) = \phi_{i}(\lambda) \cdot A \cdot t_{\text{exp}}
]
where (\phi_{i}(\lambda)) is the photon flux, (A) is the effective area of the pixel, and (t_{\text{exp}}) is the exposure time.
-
Quantum Efficiency (QE): The number of photoelectrons generated is:
[
N_{i}^{\text{pe}}(\lambda) = \eta_{i}(\lambda) \cdot P_{i}(\lambda)
]
where (\eta_{i}(\lambda)) is the quantum efficiency, which varies with wavelength and pixel.
-
Dark Current: The dark current contributes additional electrons:
[
N_{i}^{\text{dark}} = D_{i} \cdot t_{\text{exp}}
]
where (D_{i}) is the dark current per pixel.
-
Readout Noise: The readout noise is typically Gaussian:
[
N_{i}^{\text{read}} \sim \mathcal{N}(0, \sigma_{\text{read}}^2)
]
where (\sigma_{\text{read}}) is the standard deviation of the readout noise.
-
Total Signal: The total signal in digital numbers (DN) is:
[
DN_{i}(\lambda) = G_{i} \cdot \left( N_{i}^{\text{pe}}(\lambda) + N_{i}^{\text{dark}} + N_{i}^{\text{read}} \right) + O_{i} + S_{i}
]
where (G_{i}) is the gain (e⁻/DN), (O_{i}) is the optical offset (e.g., stray light), and (S_{i}) is the structured additive signal (e.g., vignette).
Calibration Pipeline
To achieve the desired accuracy, the calibration pipeline should include the following steps in the specified order:
-
Bias/Pedestal Removal:
- Subtract the bias frame (median of dark frames) to remove fixed-pattern noise and amplifier offsets.
[
DN'{i}(\lambda) = DN{i}(\lambda) - \text{Bias}_{i}
]
-
Vignette Subtraction:
- Use the "vignette" detector rows to estimate and subtract the structured additive signal.
[
V_{i}(\lambda) = f_{V,i}(\lambda) \cdot t_{\text{exp}}
]
where (f_{V,i}(\lambda)) is the vignette function per pixel.
[
DN''{i}(\lambda) = DN'{i}(\lambda) - V_{i}(\lambda)
]
-
Dark Subtraction:
- Estimate the dark current from dark frames and subtract it.
[
D_{i} = \text{Median}(DN''{i}(\lambda) \text{ for dark frames})
]
[
DN'''{i}(\lambda) = DN''{i}(\lambda) - D{i} \cdot t_{\text{exp}}
]
-
Gain Calibration:
- Determine the gain per column (not per pixel) to avoid confusion with the strong QE roll-off.
- Use uniform lamp exposures to estimate gain.
[
G_{j} = \frac{\text{Mean}(DN'''{i}(\lambda))}{\text{Known Photons}}
]
[
Q{i}(\lambda) = \frac{DN'''{i}(\lambda)}{G{j}}
]
-
Stray Light Subtraction:
- Estimate and subtract stray light using regions known to be dark or using a model based on the lamp edges.
[
S_{i}(\lambda) = \sum_{k} a_{i,k} \cdot B_{k}(\lambda)
]
where (B_{k}(\lambda)) are basis spectra and (a_{i,k}) are coefficients.
[
Q'{i}(\lambda) = Q{i}(\lambda) - S_{i}(\lambda)
]
-
Non-Linearity Correction:
- Apply a non-linearity correction to the signal.
[
Q''{i}(\lambda) = f{\text{NL}}(Q'_{i}(\lambda))
]
-
Flat-Field/QE Correction:
- Use a flat-field frame to correct for pixel-to-pixel variations and apply QE correction.
[
\phi_{i}(\lambda) = \frac{Q''{i}(\lambda)}{\eta{i}(\lambda)}
]
Impact of Omitting Gain Variation or Non-Linearity Correction
Summary Pipeline
Raw DN
↓
Bias/Pedestal Subtraction
↓
Vignette Subtraction
↓
Dark Subtraction
↓
Gain Calibration (per column)
↓
Stray Light Subtraction
↓
Non-Linearity Correction
↓
Flat-Field/QE Correction
↓
Final Photon Spectrum
By following this pipeline, you can achieve the desired systematic error of better than 0.5% in measuring accurate absorption lines.