| Primary actuator | Direct scalp current | Time‑varying magnetic field (coil) | Focused acoustic pressure wave |
| Typical stimulus magnitude | 0.5–2 mA total; electrode current density ≈0.02–0.08 mA/cm² | Coil currents ~kA; Bpeak ≈1–2 T at coil face | Frequency 0.2–1.0 MHz; peak negative pressure (PNP) at focus ≈0.1–0.8 MPa (after skull attenuation) |
| Induced electric field (cortex) | ≈0.05–1 V/m (≈0.2–0.5 V/m typical for 1–2 mA) | ≈50–300 V/m (≈100–200 V/m at motor threshold) | Ultrasound is primarily mechanical — direct induced E‑field ≪1 V/m (negligible compared with TMS/tES) |
| Spatial focality (lateral FWHM) | Conventional pads: ~5–10 cm; HD montages: ~1–3 cm | Figure‑8 coil: ~1–2 cm | Very focal: ≈1–6 mm lateral (axial 5–20 mm) depending on transducer/frequency |
| Effective depth | Superficial cortex (main effects ≤1–2 cm) | Cortex ≈1.5–3 cm (deep coils H‑coil ≈4–6 cm, less focal) | Adjustable focus — can reach deep structures (>6 cm) depending on transducer/steering |
| Temporal precision | DC for minutes; tACS: 0.1–200 Hz (slow neuromodulation) | Pulse width ~0.1–0.5 ms; 1–100+ Hz (sub‑ms control) | Pulses µs–ms; PRF up to kHz — ms‑scale temporal control |
| Neural effect | Subthreshold modulation (membrane polarizations <1 mV); modulates excitability/plasticity | Suprathreshold: directly evokes action potentials (e.g., MEPs) | Variable: can be sub‑ or suprathreshold; animal thresholds often PNP ≈0.3–0.7 MPa; human data mostly modulatory so far |
| Typical safety constraints | Common clinical/research ≤2 mA; electrode density <0.1 mA/cm² | Follow rTMS safety tables (seizure risk at high freq); intensity often referenced to % motor threshold | Avoid heating/cavitation; follow MI and intensity guidance (diagnostic limits: MI<1.9, Ispta guideline ≈720 mW/cm²); many protocols use low duty cycles |