| Principal driver | Direct or alternating current via scalp electrodes; diffuse spread | Time-varying magnetic field induces cortical electric field | Focused acoustic pressure waves via transducer; mechanical modulation |
| Typical scalp intensity / dose | 0.5–2 mA (often up to 4 mA in some studies) | Stimulation intensity set as % of Resting Motor Threshold (RMT); commonly 110–120% RMT | Pulsed ultrasound with Ispta or PRPA indices; common ranges: Ispta ~0.1–1.0 W/cm²; PRPA ~0.3–0.8 MPa |
| Electrode/transducer size or focus | Electrodes ≈ 25–35 cm² (e.g., 5×5 cm to 5×7 cm); diffuse field | Figure‑8 coils yield focal area ~8–15 mm diameter; deeper/helical coils reduce focality | Focal spot typically mm‑scale (2–5 mm) at target depth; higher‑endeavors reach several cm depending on focus |
| Peak cortical electric/magnetic field | Cortex E-field ≈ 0.2–0.5 V/m for typical 1–2 mA; skin current density ≈ 0.04–0.08 mA/cm² | Cortex E-field ≈ 80–150 V/m (depends on coil and depth); coil B‑field surface ~1–2 T | No direct E/M field like TMS; brain effect via mechanical indices; ultrasound is described by PRPA and Ispta |
| Target depth (effective) | ~1–2 cm below skull (near surface cortex) | ~1.5–2 cm; deeper with specialized coils (e.g., H‑coil) but less focal | ~2–6 cm with focused transducers; deeper targets require larger apertures or higher frequency adjustments |
| Focality (typical) | Diffuse; multiple cm² to tens of cm² depending on montage | Focal region ~1–2 cm diameter (with standard coils); deeper focus at cost of focality | Very focal (mm‑scale) to a few cm³ depending on transducer and skull window |
| Typical session length | 10–30 minutes per session | 5–20 minutes per session (single/pulse or repetitive protocols) | Pulsed protocols; total session ~1–20 minutes depending on protocol |
| Common protocols / frequencies | DC (tDCS) or AC in 1–40 Hz range (tACS); RNS/tRNS up to high kHz bands | rTMS 1–20 Hz; theta-burst (e.g., 50 Hz bursts every 5 Hz); single pulses | Pulsed ultrasound at ~0.25–0.8 MHz; duty cycles of a few percent to tens of percent; burst durations ms–s |
| Safety profile (typical) | Mild tingling, skin irritation; burns risk if poorly applied | Seizure risk extremely low with guidelines; headaches; scalp discomfort | Heating/mechanical effects; generally safe at low intensities; long‑term safety still under study |