Let's break down the statement and explain it in detail:
Statement: "The decreasing of absorbance that are monitoring at 340 nm is proportional to the activity of AST in the sample."
Components of the Statement:
-
Decreasing of Absorbance:
- Absorbance: A measure of the amount of light absorbed by a sample at a specific wavelength. It's often measured using a spectrophotometer.
- Decreasing: Indicates that the absorbance value is reducing over time.
-
Monitoring at 340 nm:
- 340 nm: A specific wavelength of light (in the ultraviolet-visible spectrum) at which the absorbance is being measured.
- Monitoring: Suggests that the absorbance at this wavelength is being continuously or intermittently measured over a period.
-
Proportional to the Activity of AST:
- Proportional: Implies a direct relationship where an increase in one quantity (in this case, the activity of AST) leads to a predictable increase in another quantity (the rate of decrease in absorbance at 340 nm).
- Activity of AST: Refers to the enzymatic activity of Aspartate Aminotransferase (AST), an enzyme found in various tissues of the body. AST catalyzes the transfer of an amino group from aspartate to α-ketoglutarate, forming glutamate and oxaloacetate.
Explanation of the Relationship:
The statement is typically related to a biochemical assay for measuring the activity of AST in a sample (e.g., blood serum, tissue extract). Here’s how it works:
- Reaction Basis: In the assay, AST catalyzes the conversion of aspartate and α-ketoglutarate into oxaloacetate and glutamate.
- NADH Consumption: Often, this reaction is coupled with a secondary reaction involving Malate Dehydrogenase (MDH), where oxaloacetate is reduced to malate by NADH, resulting in the oxidation of NADH to NAD+.
- Absorbance at 340 nm: NADH absorbs light strongly at 340 nm, whereas NAD+ does not. Therefore, as NADH is consumed (oxidized to NAD+) in the coupled reaction, the absorbance of the sample at 340 nm decreases.
- Proportionality to AST Activity: The rate of decrease in absorbance at 340 nm is directly proportional to the amount of AST enzyme present in the sample. This is because the more AST enzyme is present, the faster the primary reaction proceeds, consuming more NADH in the coupled reaction per unit of time, and thus leading to a faster decrease in absorbance at 340 nm.
Potential Sources of Error Affecting Accuracy:
While the assay is reliable, several factors can affect its accuracy:
- Temperature and pH variations: Changes in temperature or pH can alter enzyme activity, leading to inaccurate measurements.
- Reagent quality and stability: Impurities or degradation of reagents (e.g., NADH, enzymes) can impact reaction rates and absorbance readings.
- Sample handling: Inadequate sample preparation, such as hemolysis in blood samples, can interfere with the assay.
- Instrument calibration: Incorrect calibration of the spectrophotometer can result in inaccurate absorbance measurements.
In Summary: The faster the decrease in absorbance at 340 nm, the higher the activity of AST in the sample, due to the direct relationship between AST enzyme activity and the rate of NADH consumption in the coupled enzymatic reactions.