Only pressure and temperature transmitters are subjected to calibration using a standard measuring instrument, and the pressure and temperature values and the status of the communication link are constantly monitored through the flow computer.
The sensitive element of the temperature transmitter is type Pt 100, class A, according to BDS EN 60751 Industrial platinum resistance thermometers and platinum temperature sensors (IEC 60751:2008).
A volume flow rate correction method is used for offsetting the systematic errors of the ultrasonic flow meters and for maintaining the flowmeters’ measurement uncertainty within the specified one for the measurement range. The error values at the respective calibration points of the ultrasonic flow meters are introduced into the flow computers through the linear interpolation method.
The maximum permissible deviation of the corrected volume between Pay and Check system shall be no more than ±0,5 % on daily basis.
Flow computers calculate:
- Gas volume (m3) and volume flow rate (m3/h) at the pressure and temperature of the metering lines;
- Gas volume (m3) and volume flow rate (m3/h) after applying the error curve according to the respective ultrasonic flow meter’s calibration certificate at the pressure and temperature of the metering lines;
- Gas volume and volume flow rate at reference pressure conditions of 1,01325 bar(a) and temperature of 273,15 K (0°C);
- Gas compression factor according to EN ISO 12213-2 Natural gas - Calculation of compression factor - Part 2: Calculation using molar-composition analysis;
- Gross calorific value according to ISO 6976;
- Energy (MWh) and energy rate (MWh/h) based on the gross calorific value and volume of natural gas.
The following reference conditions are used:
Temperature 273,15 K (0 °C)
Pressure 1,01325 bar(а)
Combustion temperature 298,15 K (25° C)
The volume is determined at reference conditions (0 ͦC) and is converted/calculated at conditions (20 ͦC) in compliance with ISO 13443 Natural gas - Standard reference conditions.
Flow computers report the flow from the ultrasonic flow meter by using RS485 communication and a digital impulse, which is the main input for measurement. The gas flow is adjusted for nonlinearity of the ultrasonic flow meter based on the data of the flow meter’s calibration performed by an accredited laboratory with acknowledged competence. The flow computer performs interpolation between the values entered at the calibration points.
Flow computers receive data on the gas composition used by the fiscal system.
All data generated by the flow computers are the actual reported data stored and processed by the fiscal system. All the results of the calculations are stored in the respective registers of the flow computers. The data on pressure, temperature, compression factor are constantly updated to enable calculation of the average flow by the fiscal system.
The fiscal system generates all diagnostic and fiscal reports for natural gas.
For the purposes of control over the measurement, the fiscal system performs the following functions:
- sums up the quantities (energy and volume) of all metered flows;
- verifies and transmits the appropriate qualitative analysis of the chemical composition to the flow computer;
- calculates the required quantity reports (under measurement conditions and reference conditions) and compiles reports on a daily (hourly) and monthly (daily) basis;
- stores the measured and calculated (qualitatively and quantitatively) data in files
- guarantees the integrity of these files and records any changes in them;
- provides a secure/guaranteed mechanism for accessing the parameters at the metering station;
- communicates with the SCADA system, transmitting data and measurement conditions.
The files generated by the fiscal system are used in the preparation of reports for the purpose of reporting natural gas quantity and quality. Copying of the generated reports is carried out by Bulgartransgaz EAD qualified personnel with measures undertaken to prevent damage to equipment and measurements records.
4.1.2. Measuring Instruments Monitoring and Verification
The measuring instruments included in the metering systems are periodically inspected, checked and calibrated.
The inspection and calibration procedures as well as the acceptance criteria for the results are detailed in Appendix No. 3.
The frequency of calibration, verification and inspection, as well as the type of the applicable procedure shall be in accordance with those laid down in Table 1:
Table1
Element of
the metering system Procedure type Period
Ultrasonic flow meter and PTZ volume converter Check by sequentially connecting the metering line with the reference turbine flow meter and a metering line in operation 1 year
Pressure transmitter at the metering lines Check by a reference measuring instrument 1 year
Temperature transmitter at the metering lines Check by a reference metering instrument 1 year
Gas chromatograph Automatic calibration with certified reference gas Once a day
Turbine flow meter Calibration at a testing stand at high pressure 6 years
Water dew point analyser Check by a reference measuring instrument 1 year
Hydrocarbon dew point analyser Check by a reference measuring instrument 1 year
Ultrasonic flowmeter Calibration at a testing stand at high pressure 8 years
Bulgartransgaz EAD prepares an annual calibration, check and inspection plan which is submitted to TRANSPORTGAS for information purposes by the end of November of the previous year.