Deriving Wheatstone bridge balance and strain gauge use
At balance, zero galvanometer current means no voltage drop across it, so the midpoint potentials are equal; using voltage division on each branch and setting these equal yields the classic ratio condition R1/R2 = R3/R4. This ensures the bridge output voltage is zero when balanced. Strain gauges use this setup by acting as a variable resistor within the bridge—small resistance changes from strain unbalance the bridge, producing a measurable output proportional to strain.
A convenient way to think about a Wheatstone bridge is as two independent voltage dividers fed from the same source. Label the corners of the diamond A–B–C–D as shown, put (R_{1}) between A and B, (R_{2}) between B and C, (R_{3}) between A and D, (R_{4}) between D and C, feed a supply (V_{\mathrm{s}}) between A and C, and connect the galvanometer between the mid‑points B and D:
A o──R1──o B
| |
R3| |R2
| |
D o──R4──o C
With the galvanometer removed or carrying no current the two arms are independent, and the potentials at B and D are simply the divider voltages. Taking (V_C) as zero and (V_A=V_{\mathrm{s}}), the currents are
[
I_1 = \frac{V_{\mathrm{s}}}{R_1+R_2}\quad\text{through }R_1,R_2,\qquad
I_2 = \frac{V_{\mathrm{s}}}{R_3+R_4}\quad\text{through }R_3,R_4,
]
and the node voltages are
[
V_B = V_{\mathrm{s}} - I_1 R_1 = V_{\mathrm{s}}\frac{R_2}{R_1+R_2},\qquad
V_D = V_{\mathrm{s}} - I_2 R_3 = V_{\mathrm{s}}\frac{R_4}{R_3+R_4}.
]
The galvanometer current is driven by the difference (V_B-V_D). If that current is zero, there is no voltage drop across the galvanometer coil, so (V_B=V_D), which leads directly to the balance condition
[
\frac{R_1}{R_2} = \frac{R_3}{R_4}.
]
Conversely, if (R_1/R_2=R_3/R_4) then the two divider outputs are equal and the bridge is balanced at zero output.
This property is exploited in resistive strain measurements. A strain gauge is a length of fine wire or foil whose resistance (R) changes slightly when it is stretched or compressed. By incorporating one (quarter‑bridge), two (half‑bridge) or four (full‑bridge) gauges into the arms of a Wheatstone bridge and initially balancing it with equal nominal resistances, the bridge output is zero at zero strain. A small change (\Delta R) due to mechanical strain unbalances the bridge and produces a differential output voltage
[
V_{\text{out}}\approx V_{\mathrm{s}}\left(\frac{R+\Delta R}{R+R+\Delta R} - \frac{R}{R+R}\right)\propto \frac{\Delta R}{R},
]
which can be amplified and related to the applied strain via the gauge factor. The bridge configuration makes the measurement sensitive to very small resistance changes and allows for temperature compensation by appropriate choice of the other arms.