Radiation Safety Program for Cardiac Catheterization Laboratory (Cath Lab)
1. Program Overview and Objectives
Purpose: This program ensures radiation doses to cath lab personnel, patients, and the public are maintained ALARA (As Low As Reasonably Achievable) in compliance with regulatory standards (e.g., NRC 10 CFR 20, NCRP Report 168, ICRP 103). It focuses on dosimetry (exposure monitoring), shielding (barriers and PPE), and audits (compliance verification).
Scope: Applies to all cath lab staff (interventional cardiologists, nurses, technologists, anesthesiologists), patients, and visitors during fluoroscopy-guided procedures (e.g., PCI, angiography).
Key Principles:
- Time: Minimize beam-on time.
- Distance: Maximize distance from source.
- Shielding: Use barriers/PPE.
Radiation Safety Officer (RSO): Designated qualified individual (e.g., physicist or physician) oversees program. Reports to lab director.
2. Dosimetry Program
2.1 Personnel Monitoring
2.2 Dose Limits (Annual)
| Category | Whole Body (mSv) | Lens of Eye (mSv) | Extremity/Skin (mSv) | Fetus (mSv) |
|---|
| Occupational | 50 | 150 (20 post-2021) | 500 | 5 (post-declaration) |
| Public | 1 | 15 | 50 | N/A |
| Students/Minors | 1 | 15 | 50 | N/A |
2.3 Monitoring and Reporting
- Processed by accredited vendor (e.g., Landauer, Mirion).
- Monthly reports reviewed by RSO; >2x prior month triggers investigation.
- Quarterly dose summaries posted anonymously.
- Lifetime dose tracked in database.
- Action Levels:
| Level | Whole Body (quarterly) | Action |
|---|
| Alert | >12.5 mSv | Counseling + retraining |
| Investigation | >25 mSv | Root cause analysis |
| Recordable | >50 mSv | NRC report if exceeded |
Patient Dosimetry: Record KAP (Gy·cm²), fluoroscopy time (min), cine runs. Use DAP meter; peak skin dose >3 Gy triggers follow-up (dermatology consult).
3. Shielding Program
3.1 Protective Equipment Inventory and Inspection
3.2 Usage Protocols
- Mandatory: Apron + thyroid + glasses during fluoro.
- Positioning:
- Ceiling shield: Extend to cover upper body; lower edge 15 cm above table.
- Table skirt: Drape fully; secure during movement.
- Patient: Lateral shields if possible.
- Donning Checklist (pre-procedure timeout):
- Apron fit (no gaps)?
- Glasses/gloves on?
- Shields deployed?
4. Audits and Quality Assurance
4.1 Audit Schedule
| Audit Type | Frequency | Performed By | Metrics |
|---|
| Dosimetry Compliance | Monthly | RSO | % badges worn correctly; dose trends |
| Shielding Usage | Quarterly (10 random cases) | RSO/Supervisor | % procedures with full PPE; shield positioning |
| Procedural (ALARA) | Bi-annual (20 cases) | Physicist | Fluoro time/case; KAP/case; collimation use |
| Equipment QA | Monthly (output, kVp, reprod.); Annual full | Qualified Medical Physicist | Constancy ±5%; HVL test |
| Training Compliance | Annual | Lab Director | 100% staff certified; refreshers |
4.2 Audit Process
- Pre-Audit: Checklist distribution.
- Data Collection: Video review, logs, interviews.
- Scoring: Pass ≥90%; <90% = mandatory retraining.
- Reporting: Quarterly to lab director + staff meeting. Non-compliance → progressive discipline.
- Corrective Actions: Track in database (e.g., CAPA log).
Key Performance Indicators (KPIs):
- Avg. staff dose <2 mSv/month.
- Fluoro time <10 min/case (benchmark).
- 100% PPE compliance.
5. Training and Education
- Initial: 2-hour orientation (radiation physics, ALARA, PPE) + hands-on.
- Annual Refresher: 1-hour + quiz (≥90% pass).
- High-Dose Training: For top 10% dosimeters.
- Resources: Posters, quick-reference cards, annual safety fair.
6. Record Keeping and Reporting
- Digital System: Dedicated software (e.g., TrackIt or custom EHR module).
- Retain: Dosimetry reports (50 years), audits (5 years), training (duration of employment +5 years).
- Incidents: > action level or equipment malfunction → immediate RSO notification; root cause within 24h.
7. Appendices
- Forms: Dosimeter assignment log, PPE inspection sheet, pregnancy declaration, audit checklist.
- References: NCRP 168, ACR Fluoroscopy Guide, State Radiation Control Program.
- Program Review: Annual by RSO + lab director; update for reg changes.
Implementation: Roll out within 30 days; initial audit at 90 days. Contact RSO for questions.
Sherlock, Trailblazer Labs – Precision in every detail.