Hello ! Can you help me find sensors and actuators to manage a specifically ultra low pressure (in the range of 0.01 mbar, with an accuracy of 0.001 mbar)
Assistant
Okay, managing pressure in the ultra-low range of 0.01 mbar (1 Pa) with an accuracy of 0.001 mbar (0.1 Pa) is definitely in the realm of specialized high-vacuum technology. Standard pressure sensors and actuators won't work here.
Let's break down the requirements for sensors and actuators:
Based on the search results and general vacuum technology knowledge, here's a breakdown of suitable sensors and actuators for your requirements:
1. Pressure Sensors (Vacuum Gauges):
- Technology Needed: For the pressure range of 0.01 mbar (1 Pa) and the required accuracy of 0.001 mbar (0.1 Pa), the primary choice is a Capacitance Diaphragm Gauge (CDG), also known as a capacitance manometer.
- Why CDGs?
- They measure pressure directly by sensing the deflection of a diaphragm relative to a fixed electrode. This deflection changes the capacitance, which is measured electronically.
- Crucially, this measurement principle is independent of the gas type being measured[1].
- They offer high accuracy and resolution in the low to medium vacuum range, including your target pressure.
- Many CDGs are temperature-controlled (heated) to stabilize the sensor element and minimize drift due to ambient temperature changes, improving accuracy and zero stability[2].
- Key Specifications to Look For:
- Full Scale Range (F.S.): You need a gauge with a full scale that comfortably includes your target pressure but is low enough to provide good resolution. Common low-pressure ranges are 1 mbar, 0.1 mbar, etc. A 1 mbar F.S. gauge might be suitable, as 0.01 mbar would be 1% of its full scale. Check the manufacturer's accuracy specification across the range.
- Accuracy: Specified typically as a percentage of reading[2][3]. You need a gauge where this accuracy specification meets your 0.001 mbar requirement at the 0.01 mbar point. For example, a gauge with 0.15% or 0.2% of reading accuracy would be well within your needs if its measurement range extends low enough. Some are specified as % of Full Scale, which is less ideal at the low end of the range. Be mindful of the difference between accuracy (how close to the true value) and resolution (smallest detectable change).
- Resolution: The smallest change the gauge can detect/report. This needs to be significantly better than your required accuracy (e.g., 0.0001 mbar or better).
- Temperature Control: Heated gauges (often at 45°C or higher) offer better stability.
- Materials: Ensure compatibility with any gases used (often Alumina ceramic sensors with stainless steel or Nickel alloy bodies).
- Manufacturers: Look for products from established vacuum component suppliers like:
- MKS Instruments (Baratron® gauges)
- INFICON
- Pfeiffer Vacuum
- Brooks Instrument
- Setra Systems
- VAT (though more known for valves, they may integrate or offer gauges)
2. Actuators (For Pressure Control):
To manage the pressure, you need a way to adjust it[4][5]. This typically involves either precisely letting gas into the vacuum system or controlling the rate at which gas is pumped out[5][6].
-
Controlling Gas Inflow (Increasing/Maintaining Pressure):
- Piezoelectric Leak Valves (Gas Dosing Valves): These are excellent for your requirement. They use a piezoelectric stack that expands/contracts with applied voltage, precisely opening or closing a valve orifice.
- Advantages: Very fast response (milliseconds), extremely fine control over very small gas flows, UHV compatible (often all-metal construction), fail-safe (close on power loss). Ideal for precise pressure adjustments or maintaining a setpoint via a PID controller[5].
- Manufacturers: Pfeiffer Vacuum, MKS Instruments, VAT, Oxford Applied Research, Maxtek Inc.
- Mass Flow Controllers (MFCs): While often used for higher flows, low-flow MFCs can precisely control the rate of gas flow into the system. Pressure is then controlled indirectly by balancing this inflow against the pumping speed.
- Advantages: Directly controls flow rate, good repeatability.
- Disadvantages: Slower response time compared to piezo valves, may not offer fine enough control for 0.001 mbar adjustments depending on system volume and pumping speed.
- Manufacturers: MKS Instruments, Brooks Instrument, Bronkhorst, Alicat Scientific.
- Piezoelectric Leak Valves (Gas Dosing Valves): These are excellent for your requirement. They use a piezoelectric stack that expands/contracts with applied voltage, precisely opening or closing a valve orifice.
-
Controlling Gas Outflow (Decreasing/Maintaining Pressure):
- Throttle Valves (Variable Conductance Valves): These are placed between the vacuum chamber and the high-vacuum pump. By adjusting the valve opening (often a butterfly plate or counter-rotating vanes), they restrict the pumping speed, allowing pressure control.
- Advantages: Effective for controlling pressure when there's a constant gas load or during specific process steps. Often used with PID controllers referencing a CDG. Some offer linear control characteristics[7][8].
- Disadvantages: Controls removal rate, not direct pressure addition. Requires a stable pumping system.
- Manufacturers: MKS Instruments, VAT Vakuumventile, Pfeiffer Vacuum, MeiVac (Ferrotec), Vacuum Research.
- Throttle Valves (Variable Conductance Valves): These are placed between the vacuum chamber and the high-vacuum pump. By adjusting the valve opening (often a butterfly plate or counter-rotating vanes), they restrict the pumping speed, allowing pressure control.
3. Integration and Control:
- You will need a control system (often a PID controller) that reads the pressure from the CDG sensor and adjusts the actuator (piezo valve or throttle valve) to maintain the desired pressure setpoint (0.01 mbar).
- This controller can be a standalone unit, integrated into a vacuum gauge controller, or implemented within a larger PLC or computer-based system.
Important Considerations:
- Vacuum System: The overall vacuum system design (chamber volume, pump type, pumping speed, leak rate) heavily influences controllability.
- Connections: Ensure compatibility (e.g., KF, CF, VCR flanges).
- Materials: Must be vacuum compatible and suitable for your process gases.
- Leak Tightness: Critical at these low pressures.
- Calibration: Regular calibration of the CDG is essential for maintaining accuracy.
In summary, for measuring 0.01 mbar ± 0.001 mbar, a high-accuracy, temperature-controlled Capacitance Diaphragm Gauge is the sensor of choice. For control, a Piezoelectric Leak Valve offers the finest adjustment for gas inlet, while a Throttle Valve can be used to control pumping speed. These components need to be integrated with a suitable PID control loop.