    Consider you have manufacturer of medical devices and your organization effected with serious incidents involving devices. You need to submit report/notify the EU Medical Device Regulation. So search for respective report or notification form

  2. Assistant

    As a manufacturer of medical devices, it's imperative to report any serious incidents involving your products to the competent authorities in the European Union (EU) in accordance with the Medical Device Regulation (MDR) (EU) 2017/745. The standardized form for such reporting is the Manufacturer Incident Report (MIR).

    Accessing the MIR Form:

    The MIR form is a common European reporting template designed to streamline the reporting process across EU member states. You can download the latest version of the MIR form from the Danish Medicines Agency's website:

    Guidance on Completing the MIR Form:

    To assist in accurately filling out the MIR form, refer to the "Manufacturer Incident Report Helptext 2020," which provides detailed instructions and clarifications:

    Submission of the MIR Form:

    Once completed, the MIR form should be submitted to the competent authority of the EU member state where the incident occurred. Each member state has specific contact details and procedures for submission. For example:

    • Denmark: Submit to the Danish Medicines Agency via email or postal mail. Contact details are available on their website.

    • Norway: Submit to the Norwegian Medical Products Agency (NoMA) at [EMAIL_ADDRESS]. Both PDF and XML versions of the report must be attached[1].

    • Netherlands: Submit to the Health and Youth Care Inspectorate (IGJ) at [EMAIL_ADDRESS]. Include the entire form as a PDF file and any relevant attachments[2].

    Ensure you adhere to the specific reporting deadlines stipulated by the MDR:

    • Serious public health threat: Report immediately, and no later than 2 days after becoming aware of the incident.

    • Death or unanticipated serious deterioration in health: Report no later than 10 days after becoming aware of the incident.

    • Other serious incidents: Report no later than 15 days after becoming aware of the incident.

    For comprehensive information on reporting procedures, deadlines, and additional guidance, consult the relevant competent authority's website in the country where the incident occurred.