Preservation of Samples for ICP (Inductively Coupled Plasma) Analysis and the Use of Nitric Acid
1. Preservation of Samples for ICP
When preparing samples for ICP analysis (such as ICP-OES or ICP-MS), proper preservation is crucial to ensure accurate and reliable results. The main goals of sample preservation are:
- Preventing contamination (from the environment or containers)
- Preventing loss of analytes (due to precipitation, adsorption, or volatilization)
- Maintaining the chemical integrity of the sample until analysis
2. Use of Nitric Acid for Storage
Nitric acid (HNO₃) is commonly used for preserving samples for ICP analysis, especially for water, soil, and biological samples. The typical procedure involves acidifying the sample to a pH < 2 by adding concentrated nitric acid (usually 1–2% v/v final concentration).
Reasons for Using Nitric Acid:
-
Prevents Precipitation of Metals:
- Many metal ions can form insoluble hydroxides or carbonates at neutral or basic pH. Acidifying with nitric acid keeps metals in solution by preventing the formation of these precipitates.
-
Minimizes Adsorption to Container Walls:
- Acidic conditions reduce the tendency of metal ions to adsorb onto the walls of glass or plastic containers, which could otherwise lead to underestimation of their concentrations.
-
Inhibits Microbial Activity:
- Acidic conditions inhibit the growth of microorganisms that could alter the chemical composition of the sample by metabolizing or transforming analytes.
-
Oxidizes Organic Matter:
- Nitric acid is a strong oxidizer and can help break down organic matter, releasing metals that may be bound to organic molecules, thus ensuring total metal content is measured.
-
Prevents Volatilization of Some Elements:
- Acidification can help prevent the loss of volatile metal species (e.g., mercury, arsenic) by converting them to less volatile ionic forms.
Summary Table
| Reason | Effect of Nitric Acid |
|---|
| Prevents precipitation | Keeps metals dissolved |
| Minimizes adsorption | Reduces loss to container walls |
| Inhibits microbial activity | Stops biological changes |
| Oxidizes organic matter | Releases bound metals |
| Prevents volatilization | Stabilizes volatile elements |
References
- US EPA Method 200.7: "Determination of Metals and Trace Elements in Water and Wastes by ICP-AES"
- Standard Methods for the Examination of Water and Wastewater
In summary:
Samples for ICP analysis are preserved with nitric acid to keep metals dissolved, prevent contamination or loss, and maintain sample integrity until analysis.