I'd like to validate a new measurement technique. What statistical method(s) are appropriate to ensure that the value of the new technique is not different from the value of the old technique?
Assistant
Other models, same or similar prompt
Responses were produced in different, uncontrolled contexts (system prompts, settings, prior turns). This is not a benchmark.
When you want to show that a new measurement technique gives “the same” values as an established one, you are NOT trying to prove “no difference” with a classical null-hypothesis test (e.g. paired t-test).
Failing to find a significant difference only means “we don’t have evidence of a difference,” not “they are equal.”
Instead, you need a suite of agreement / equivalence tools built around three complementary questions:
Do the two techniques give the same average value (no constant bias)?
Is there no proportional bias across the measurement range?
Is the random disagreement small enough to be acceptable for practical use?
The standard toolbox looks like this:
Goal
Recommended statistic
What you conclude from it
Equivalence of means (constant bias)
Two One-Sided Tests (TOST) for equivalence, or confidence interval of mean difference
The average difference lies inside a pre-specified equivalence margin ±Δ
Absence of proportional bias
Deming or Passing-Bablok regression (both-sided errors)
Slope ≈ 1 within tolerance; intercept ≈ 0
Overall agreement (random error)
Bland-Altman 95 % Limits of Agreement, Intraclass Correlation Coefficient (ICC)
STEP 0 Plan before you measure
• Pick an equivalence margin Δ that is practically/clinically acceptable (e.g. ±2 mm Hg for BP, ±0.5 °C for temp, ±3 % for concentration, …).
• Power : do a sample-size calculation for an equivalence test (you usually need more subjects than for a difference test).
STEP 1 Collect paired data
Measure each sample (or subject) with BOTH methods. Keep raw pairs (Oldᵢ, Newᵢ).
If both one-sided p-values < α (or the 90 % CI of the mean difference lies inside ±Δ), the means are equivalent.
STEP 3 Check proportional bias
Fit Deming regression (errors in both X and Y) or the non-parametric Passing-Bablok regression:
Slope ≈ 1 and intercept ≈ 0 (within tolerance) ⇒ no proportional or constant bias.
Visual: add the 45° identity line and the regression line.
STEP 4 Quantify random disagreement
Bland-Altman:
Differences dᵢ = Newᵢ – Oldᵢ
Averages aᵢ = (Newᵢ + Oldᵢ)/2
Plot dᵢ vs aᵢ
Limits of Agreement = d̄ ± 1.96·SD(d)
Check whether those limits fall inside the domain you consider acceptable (often ±2·Δ or a domain dictated by regulation).
Calculate the Intraclass Correlation Coefficient (ICC, preferably ICC(2,1) or ICC(A,1)). ICC ≥ 0.90 is usually “excellent” agreement.
STEP 5 Report everything transparently
• The predetermined Δ and why it is clinically / technically justified.
• Sample size and power.
• TOST p-values or CI, regression slope & intercept with CI, LoA and ICC.
• Graphs: Bland-Altman plot with LoA, scatterplot with identity + regression.
Why not just a paired t-test or correlation?
• A paired t-test asks “is the mean difference non-zero?”—you can never prove zero.
• High correlation (R²) only says the two methods vary together. Two scales that differ by a constant offset or slope can still correlate at 0.99 but be clinically useless.
Key references
• CLSI guideline EP09-C: “Measurement Procedure Comparison and Bias Estimation Using Patient Samples.”
• Bland & Altman (1986). The famous paper introducing LoA.
• Schuirmann, D. (1987). The origin of TOST.
• ISO 5725-1 to 6: Accuracy of measurement methods and results.
Take-away
Define an acceptable margin Δ, use an equivalence test (TOST) for the mean, complement it with regression for bias and Bland-Altman / ICC for random agreement. Combine the evidence to decide if the new technique can replace the old one.