Radiology Lec. 3 X – Ray Films د. ﺷﮭﺪ ﺿﺮﻏﺎم Radiograph: Is the image of an object made…
- User
Radiology
Lec. 3 X – Ray Films د. ﺷﮭﺪ ﺿﺮﻏﺎم
Radiograph: Is the image of an object made with use of X- ray instead of light.
Dental x- ray film: Is a recording media on which image of the object was made by exposing this film to X- ray.
Types of X- ray film
a- Intra oral X- ray film. b- Extra oral X- ray film.
a- Intra oral X- ray film
Chemical composition of X- ray film:
It consists of a sensitized emulsion present on both sides of transparent base.
The base is the foundation of the radiographic film , made from cellulose acetate. Its primary purpose is to provide a rigid structure onto which the emulsion can be coated. Its flexible and fracture resistant to allow easy handling but rigid enough to be placed on the viewer.
The emulsion is the heart of the x-ray film , it’s the material with which the x-ray or light photons interact and forming the image. It consists of homogenous mixture of silver halides crystals (mainly silver bromides) suspended in gelatin. The silver bromide crystals are sensitive to both light and X- ray photons.
Fig. 1: Diagram showing the cross-sectional structure of radiographic film.
The intra oral film is wrapped by opaque material to prevent light from reaching the film because light photons can activate the silver halides crystals. Also a thin sheet of
(Lead foil) is usually placed behind the film to prevent most of secondary radiation that originated in the tissue of the patient behind the film from reaching it.
Therefore this lead foil reduces secondary radiation and minimizes film fog. In addition the lead foil absorbs X- ray that have passed through the object and the film so it reduce the exposure of the tissue behind the film. This foil has a design of (herring bone pattern).
Intra oral film types
Fig. 2: The contents of a film packet.
Classified on numerical basis into:
A - Type I Called periapical film used to examine the apical area of the tooth and the surrounding structures (record the crowns, roots, and surrounding bone). Film packs come in three sizes •Size 0 for small children (22 mm × 35 mm) •Size 1, which is relatively narrow and used for views of the anterior teeth (24 mm × 40 mm) •Size 2, the standard film size used for adults (30.5 mm 40.5×mm) So the available sizes are (1.0, 1.1 and 1.2)
B – Type II Is called bitewing film it used to detect the inter proximal caries and the height of alveolar bone between 2 adjacent teeth. Bite-wing films often have a paper tab projecting from the middle of the film on which the patient bites to support the film. Size 2 film is normally used in adults; the smaller size 1 is preferred in children. In small children, size 0 may be used. A relatively long size 3 is also available. So the size include (2.0, 2.1, 2.2 and 2.3)
C – Type III Is called occlusal film that used to demonstrate area larger in dimension than area appearing in periapical film. the size is (3.4) only.
Self-developing type
Fig 4: types of x-ray films (manual and self-developing)
Intra oral film speed
Speed means the sensitivity of X- ray film silver bromide crystals (Ag Br) to X- ray photon.
There is direct relation between the speed of the film and the size of the crystals, the larger crystal size the faster film speed. the faster mean it need less amount of radiation to produce radiographic image so less radiation dose absorbed by patient.
The classification of film speed based on alphabetical basis so from A to F, film speed A is the slowest while speed F is the faster one.
b- Extra oral film The purpose of using such film is to make a radiographic image able to examine an area in and around the jaw that can't be seen by intra oral film, such as panoramic , cephalometric and other skull radiograph
Types of extra oral film
- Screen films
- Non screen films
Non screen film
- Film emulsion is more sensitive to X- ray than to light.
- The film has double emulsion like intra oral film but the emulsion is thicker.
- Increased thickness of emulsion make the non screen film need less amount of radiation so it need less exposure time.
- The size of the film used include: 5×7 and 8×10 inches.
Screen film
- Film emulsion is more sensitive to visible light and more specifically to blue light in the visible light spectrum.
- The size include: - 5×7, 8×10 and 10×12 inches. The screen film placed between 2 fluorescent screen in cassette. These 2 fluorescent screen made from (tiny calcium tungestate crystals). When these crystals exposed to X- Ray , the result of this exposure is a creation of light , this light in turn exposes the screen film to produce the image.
Fig4: screen film
Film properties These include density, contrast and details or definition. A. Density: Is the degree of blackness present in the processed film it measures in terms of light transmission on a percentage or logarithmic scale. Film density used in diagnostic radiograph is ranged from 0.25 to 2.
Sensitometry is the study of the relationship between the intensity of exposure of the film and the blackness after film processing.
- Exposure time: increase exposure time increases the film density.
- Milliampere: increase milliampere value (mA) which is usually ranged from 10 – 15mA, cause increasing film density.
- Kiolvoltage: increase Kilovoltage value (kV) cause increasing film density.
- Developing time: developing time usually range from 4 – 5 minutes. increase developing time cause increasing film density.
- Distance: increase the distance between x- ray tube and the film during exposure cause decrease film density.
B. Contrast: It means the graduation of differences in film density at different areas of a radiograph.
Type of contrast:
- Long – scale or low contrast:- when many different film densities can be seen between totally clear and totally black areas of the radiograph.
- Short – scale or high contrast:- when few different film densities can be seen between totally clear and totally black areas of the radiograph.
The stepwedge or penetrometer: Is an object used to show the radiographic contrast .it's usually made of aluminum and is constructed so that there is a constant increase in thickness of aluminum between the X- ray tube and the film.
Factors affect contrast
- Kilovoltage: increase kilovoltage cause increase the contrast scale
- Processing solution temperature: increase the temperature cause decrease of contrast scale.
Factors affect film density
Factors affect details
- Focal spot size: size of focal spot must be as small as possible in order to produce sharp image.
- Film grain size (film crystals): increase the size of film grain produce less sharp image.
- Movement of patient head or X- ray tube or the film during exposure causes unsharp image.
- Target object distance: which should be as great as possible, otherwise the image will be unsharp.
- Object film distance: should be as small as possible to produce sharp image.
- Screen – film contacts: poor contact cause un sharp image.
What will happen during exposure of X- ray film exposure to radiation?
x- ray photons interact with electrons of the atoms of the chemical emulsion in the X- ray film so the result is analog image, analog means the image appears identical to the original.
Latent image formation
The Ag Br crystals in the film emulsion are changed whenever they absorb X- ray photons, the result of absorption is precipitation of speck of silver in each exposed Ag Br crystal to X- ray, collectively these specks are called Latent image which is invisible and in order to convert to visible image X- ray film must be processed.
Film processing Its either manual or automatic processing.
Processing cycle include: Developing, rising, fixing, washing and drying.
Developing: is the stage of processing during which the latent image is converted to a visible image.
X- ray film is placed in alkaline developer solution ,the action of developing agents are on exposed Ag Br crystals to continue the process of precipitating the specks of silver until all silver is deposit at the site of crystal and the bromine is released into the developing solution causing softening of the X- ray film emulsion .
Rinsing: by water for 30s to terminates the developer action and remove chemicals from emulsion.
C. Details or definition: Is the ability to reproduce sharp outlines of the object.
Fixing: by using a fixer solution .its action is:
- Re harden the film emulsion
- Removed all the unexposed or undeveloped crystals.
After fixing the film washed in running water & finally drying.
Dark Room
The darkroom or processing room is a place where the necessary handling and processing of radiographic films can be carried out safely and efficiently without hazard of producing film fog by accidental exposure to light or x-ray. It may exclude all outside light and provides the artificial safelight only.
Size and location of darkroom
Whenever possible oral radiography darkroom should be designed when the dental office is planned and should be convenient and easy to work with.
The size of the darkroom depends on the followings:
1.Type & amount of the films to be processed , the greater workload need larger darkroom . Large films need large processing tanks , so it takes more space in the darkroom.
2.Extra space must be provided if more than one person works with ,9sq.ft. for one person is enough but it is advisable to have at least 20 sq. ft. of floor space for average dental office.
While for the location of darkroom, many requirements should be taken in consideration:
- It can be conveniently reached from the rooms where the films exposed & examined.
- Darkroom should be located where room temperature fluctuates as little as possible because the temperature of the processing solution must kept constant .
- Humidity retards drying of the processed films and damages unused films stored in opened films boxes.
- The darkroom should be accessible to plumbing & power lines.
- The darkroom must also be well ventilated to provide a comfortable working environment
Illumination of dark room
-
A ceiling light to provide ordinary illumination in the darkroom , its switch must be placed high enough on the wall to prevent the operator from accidentally turning it on during processing .
-
Safe light , it consist of a filtered light beam .
-
Red warning light which is placed outside the entrance to the room , it should be wired so that it is illuminated whenever the safelight is turned on
Film storage
1.Film must be stored away from excessive heat and humidity. 2.Chemicals must not be allowed to come in contact with stored films. 3.Objects should not be placed on top of stored films because pressure can cause film artifacts. 4.The boxes of stored films should be lead lined or made of steel to prevent stray radiation from fogging the films.
Fig. 5: typical characteristic curve of indirect-action radiographic film, showing the main regions of the curve including background fog density, toe and shoulder.
Fig. 6: characteristic curve of radiographic film is the graphic relationship between optical density (OD) and exposure.
Fig. 7: comparison of speed between insight and ultra-speed film. insight film is faster .
Fig. 8: comparison of contrast between film and B. film A is higher contrast than film (straight-line portion is greater for A).
Ref. White and Pharaoh’s. Oral radiology principles and interpretation. 8th edition 2019.
make 10 quastion with false and true answer 7 of them hard and write there answers
- Assistant
-
True or False: Dental x-ray film is a recording media that captures images using light instead of X-rays. False. Dental x-ray film captures images using X-rays instead of light.
-
True or False: The base of an x-ray film is made from cellulose acetate, which provides a rigid structure for the film. True.
-
True or False: The emulsion in x-ray film consists of a mixture of silver halide crystals suspended in gelatin, which are sensitive to both light and X-ray photons. True.
-
True or False: Lead foil is placed behind the x-ray film to prevent light from reaching the film and activating the silver halide crystals. False. Lead foil is placed behind the film to prevent most of the secondary radiation from reaching the film.
-
True or False: Periapical film is used to examine the apical area of the tooth and surrounding structures, and it comes in three different sizes. True.
-
True or False: Bitewing film is used to detect interproximal caries and the height of alveolar bone between two adjacent teeth. True.
-
True or False: Occlusal film is used to demonstrate areas larger in dimension than those seen in a periapical film. True.
-
True or False: Intraoral films are classified based on their numerical sizes, with Type II films being narrower and used for views of the anterior teeth. False. Type II films are bitewing films, which are used for interproximal caries detection and height of alveolar bone.
-
True or False: Screen films are more sensitive to X-rays than to light, while non-screen films are more sensitive to visible light. False. Non-screen films are more sensitive to X-rays than to light, while screen films are more sensitive to visible light.
-
True or False: Density refers to the degree of blackness present in the processed film, while contrast refers to the sharpness of outlines in the image. True.
-
- User
very hard 5 quastions
- Assistant
-
True or False: Increasing the exposure time during an x-ray film exposure will decrease the film density. False. Increasing the exposure time during an x-ray film exposure will increase the film density.
-
True or False: A larger film grain size produces a sharper image on the x-ray film. False. A larger film grain size produces a less sharp image on the x-ray film.
-
True or False: Poor contact between the screen and film will improve the sharpness of the image. False. Poor contact between the screen and film will result in an unsharp image.
-
True or False: Increasing the distance between the x-ray tube and the film during exposure will increase the film density. False. Increasing the distance between the x-ray tube and the film during exposure will decrease the film density.
-
True or False: The size of the focal spot should be as large as possible to produce a sharp image. False. The size of the focal spot should be as small as possible to produce a sharp image.
-
- User
another 5 very hard
- Assistant
-
True or False: Increasing the temperature of the processing solution will increase the contrast scale of the radiographic image. True. Increasing the temperature of the processing solution will decrease the contrast scale of the radiographic image.
-
True or False: Using a higher kilovoltage (kV) setting will decrease the contrast scale of the radiographic image. False. Using a higher kilovoltage (kV) setting will increase the contrast scale of the radiographic image.
-
True or False: Increasing the object-film distance will result in a sharper image. False. Increasing the object-film distance will result in a less sharp image.
-
True or False: Unsharp images can be caused by movement of the patient's head, the x-ray tube, or the film during exposure. True. Movement of any of these factors during exposure can result in an unsharp image.
-
True or False: The purpose of the fixer solution in film processing is to remove the exposed and unexposed silver bromide crystals from the emulsion. True. The fixer solution removes the unexposed silver bromide crystals and hardens the film emulsion.
-