The Electrically Evoked Compound Action Potential (ECAP) waveform provides valuable…
- User
The Electrically Evoked Compound Action Potential (ECAP) waveform provides valuable information about the auditory nerve's response to electrical stimulation for CI users. We developed and validated a fast mode (FM) by modifying the algorithm used for a traditional mode (TM) to speed up the ECAP measurement. ECAPs were compared between FM and TM with same stimulation parameters in CS-20A CI users at five electrodes, with four stimulation levels at each electrode. (1) Development: In FM, the regular forward-masking subtraction method has been modified, and the measurement sequence is optimized to increase both the speed and the reliability of the ECAP measurement. Instead of running four steps of the forward-masking subtraction method one by one with N times repeating, the new algorithm excutes N times Step A first and then N times Step B, Step C and Step D, respectively. Using this technique, the internal accumulative register can be easily achieved in the implant to store the data and deliver the results only when the set test times have been reached. To further reduce the measurement time, Step D is only processed in the first stimulation level at each electrode, and the result is used repetively for the rest of stimulation levels at this electrode. As a result, the amount of data that needs to be transferred from the implant is significantly reduced, leading to notable improvements in data transmission speed and robustness.
- Assistant
(2) Validation: The FM was validated by comparing the ECAP measurements obtained using FM and TM in CS-20A CI users. The same stimulation parameters were used in both modes, and ECAPs were collected at five electrodes with four stimulation levels at each electrode. The waveform characteristics of ECAPs, including amplitude and latency, were analyzed and compared between FM and TM.
The results showed that the FM provided comparable ECAP measurements to the TM, with no significant differences in waveform characteristics. The amplitudes and latencies of ECAPs obtained from FM were highly correlated with those obtained from TM, indicating the reliability of the FM in measuring the auditory nerve's response to electrical stimulation.
Furthermore, the FM demonstrated a significant reduction in measurement time compared to TM. This speed improvement was achieved by optimizing the measurement sequence and reducing the amount of data that needs to be transferred from the implant. The internal accumulative register in the implant allowed for data storage and delivery after the set test times were reached, further enhancing the speed and efficiency of the ECAP measurement.
- User
Could you polish what I typed before?
- Assistant
- User
However, some challenges and limitations exist, such as relatively low efficiency, potentially long collection and processing times, and uncertainty caused by noise and residual interference in the data. For example, Garcia noted that collecting ECAP for all combinations of maskers and probe electrodes took approximately 45 minutes (Garcia et al., 2023). Multiple repeated measurements and lengthy data collection are required to obtain meaningful and reliable results. This increases the time cost for research and clinical applications and may also affect the patient's comfort and cooperation. Some modified techniques have been applied to reduce artifacts and obtain more straightforward peaks(Miller et al., 2000). Still, the issue of numerous iterations, lengthy measurement times, and slow speed has not yet been resolved. The technique used for removing electrical artifact contaminations in ECAP recordings by Adel et al. (2017) was the alternating polarity technique. In comparison, the modified forward masking method was used in Dhuldhoya (2013) and Mussoi and Brown (2019). The difference in artifact-removal techniques has been shown to affect the recorded ECAP traces (Baudhuin et al., 2016). Therefore, a new algorithm with a short time and low artifact interference needed to be explored urgently.
- Assistant
However, there are some notable challenges and limitations associated with the current ECAP measurement techniques. These include relatively low efficiency, potentially long collection and processing times, and uncertainties arising from noise and residual interference in the data. For instance, Garcia et al. (2023) reported that collecting ECAPs for all combinations of maskers and probe electrodes took approximately 45 minutes, highlighting the time-consuming nature of the process. Multiple repetitions and extensive data collection are often required to ensure reliable and meaningful results, prolonging the time and effort required for research and clinical applications. Moreover, these extended measurement times can impact patient comfort and cooperation.
- User
In FM, the algorithm of measurement sequence has been modified. Step A is initially executed NA times, followed by step B, step C, and step D, which occur NB, NC, and ND times, respectively. This is made possible by the accumulative register, which can store the data and deliver the results only when the set test times have been reached. As a result, the amount of data that needs to be transferred from the implant is significantly reduced, leading to notable improvements in data transmission speed and robustness. The current FM involved sequential completion and immediate output of four stimulus types, namely ABCD. Our research explored the FM that entails the completion of NA, NB, NC, and ND, followed by N repetitions and internal chip addition, aggregating the results for output. Calculating the Response Probe as (NA - NB +NC-ND) allows for extracting the neural signal triggered by the Probe, with the average number of repetitions being N = 10.
- Assistant
In the FM, we have made modifications to the algorithm of the measurement sequence to improve efficiency and speed. The modified sequence involves executing Step A N_A times, followed by Step B N_B times, Step C N_C times, and Step D N_D times. This is facilitated by the use of an accumulative register within the implant, which allows for storing the data and delivering the results only after the set test times have been reached.
By implementing this modified sequence, the amount of data that needs to be transferred from the implant is significantly reduced. This reduction in data transfer leads to noticeable improvements in both the speed and robustness of data transmission.
Specifically, in our research, we explored an FM approach that involves completing N_A iterations of Step A, N_B iterations of Step B, N_C iterations of Step C, and N_D iterations of Step D. These iterations are then followed by N repetitions and internal chip addition, which aggregates the results for output.
To extract the neural signal triggered by the Probe, we calculate the Response Probe as (N_A - N_B + N_C - N_D). In our research, we used an average number of repetitions, N, that is equal to 10.